Few users of microscopes have an understanding of the simple, but necessary, techniques needed for accurate measurement.
Generally, texts on microscopy do not explain how to calibrate and/or measure with a microscope. Errors in magnification and dimensions are easily committed and true magnification is often omitted entirely or only approximated.
The following information will be helpful:
It is essential to understand how the eyepiece reticle can assist you. Graduations on the reticle scale have an absolute value. The absolute value is the actual size of the graduations as measured against an absolute scale.
As the reticle is viewed through the microscope eyepiece, the graduations are magnified by the power of the eyepiece. The reticle pattern is in the position of the object plane. It is at this area that the objective lens forms an image of the specimen. The reticle pattern is then superimposed on the magnified image of the specimen.
Two things are important when selecting the eyepiece power for use with a reticle:
- The total magnification desired to accurately view the specimen to be measured.
- A magnification high enough to clearly distinguish each division of the eyepiece reticle.
It is essential to understand that the magnification of the eyepiece has no relationship to the measurement capability of the reticle. It serves to magnify the reticle so that it can be seen clearly, and also serves as one factor in the total magnification of the microscope. For example: A microscope with a 10x eyepiece and a 3x objective has a total magnification of 30x. The image at the reticle is 3x and then magnified by a factor of 10x.
When a reticle is used in the microscope, an enlarged image of the specimen is projected on the reticle scale. The amount by which the specimen image is enlarged is equal to the magnification of the objective. The size of the specimen is found by dividing the absolute value of the reticle by the objective power.
To make the measuring process easier, many reticles supplied for microscopes are not of the absolute type, but rather the graduations have been accurately enlarged by a certain factor. For example, a reticle of 0.005" specifically designed for use in a measuring microscope for tool room use, with 3x objective, would have an absolute value of 0.015".
Using the methods so far discussed, accurate measurements may be taken only when the objectives can be adjusted to an exact magnification. Objectives furnished with the Nikon Measurescope are of this type and allow extremely accurate measurements without the use of a stage micrometer.
Stereo microscopes and Research microscopes use objectives which are not adjustable to a precise magnification factor. (The exception is a Zoom Stereo microscope which can be adjusted to exact objective magnification by using an eyepiece reticle and stage micrometer of known value.) While each of these objectives has a stated nominal value, they are not primarily designed for measuring applications. When accurate measurements are to be taken with these instruments, observe the following:
- Place a stage micrometer with a known value on the stage with the microscope objective in place.
- The image of the stage micrometer is superimposed on the eyepiece reticle.
- If the graduations coincide exactly, proceed with the measurements, knowing that each graduation is equal to the stated value of the stage micrometer.
- In most cases you will find that the graduations on the reticle and stage micrometer do not coincide. They may appear slightly under or slightly over the graduations of the reticle.
In order to determine the exact value of each graduation of the reticle, divide the number of graduations of the stage micrometer by the number of graduations of the reticle. Multiply the answer by the value of the stage micrometer. Example:
We have a stage micrometer with a known value of 0.005" per increment, and a reticle that takes 10 graduations to reach 9 graduations of the stage micrometer. The value of the reticle must, therefore, be lowered so that it can measure accurately.
(Stage Micrometer Graduations / Reticle Graduations) X 0.005"
9/10=0.9 x 0.005" = 0.0045"
The result of 0.9 is known as the calibration constant. The result 0.0045 is the measuring value of this particular reticle used with this particular objective.
In this example, the approximate value of the reticle was known. If the approximate value is not known, you may use the following method:
Let’s say that you have a reticle that has 12 lines coinciding with 5 lines on the stage micrometer.
Stage micrometer value = 0.005" x 5 = 0.025"
Eyepiece Reticle = 0.025" / 12 = 0.0021"
The value of each graduation of eyepiece reticle is 0.0021" when used with this objective only.
Once the value of the reticle is known, there are two methods that can be used to make accurate measurements with the eyepiece reticle. (For the exercise, let’s assume our specimen covers exactly 12 graduations.)
1. Multiply the calibrated value of the reticle by the number of graduations covered by the specimen.
0.0045" x 12 = 0.054", or
2. Uncalibrated value of the reticle multiplied by the number of graduations covered by the specimen multiplied by the calibration constant.
0.005" x 12 = 0.06 x 0.9 = 0.054"
Once the calibration constant or calibrated value of a reticle is determined, the stage micrometer need not be used for further measurements. Since the objectives are subject to plus or minus tolerances, this type of calibration must be carried out with each objective used with the eyepiece reticle.
When making precise measurements using a Stereo Zoom microscope, it will be necessary to use a stage micrometer each time the objective zooming ring is moved. Although the zoom ring is graduated with the nominal objective powers, it is impossible to return the zoom control to exactly the same position that would be necessary for precise measuring purposes.
MICROSCOPE EYEPIECES
A microscope eyepiece is designed to further magnify the primary image formed by the microscope objective, and also limit the field of view.
In Nikon’s revolutionary new CF Optical System, chromatic aberration has been corrected in both the objectives and the eyepieces independently. The result is a dramatic reduction of chromatic aberration across the entire field of view. The orange coloring around the fringe of the field which was noticeable in the compensating system has been virtually eliminated. Eyepieces of this type, when used with a reticle, exhibit no color fringing; therefore, 100 percent of the field can be utilized.
There are a number of other conventional-type eyepieces available (the Huygenian and the Ramsden being the most common). Many microscopes still offer Huygenian-type eyepieces. However, when eyepiece reticles are used, a Ramsden or Kellner eyepiece should not be used.
The Huygenian eyepiece is of simple construction, consisting of two plano-convex lenses mounted with convex sides towards the objective. The lens nearest the eye is known as the field lens. This type of eyepiece is also uncorrected, giving a blue fringe to the edge of the field and is best suited for low-power achromatic objectives. The principal image is formed between the two lenses, making it inconvenient for use with a reticle whose accuracy is affected by the aberrations of the eye lens alone.
The Ramsden eyepiece is of construction similar to the Huygenian, except that the field lens has the plane side nearest the objective. The diaphragm is located below the lens system. This type of eyepiece has the advantage of imparting less distortion to scales and lines than the Huygenian; and therefore, its main use is for micrometry, as the reticle is placed on the field diaphragm.
The Kellner eyepiece is an improved Ramsden ocular with an achromatic doublet for the eye lens, allowing the chromatic aberration of the field lens to be more fully corrected. The eyepiece has a high eye point, useful to spectacle wearers, but does suffer from some distortion. As the lower focal plane is below the field lens, any aberrations will effect the primary image and eyepiece reticle equally. The main use of the Kellner eyepiece, therefore, is in measurements with the microscope.
The compensating eyepiece is generally constructed of two separate lenses, one or both of which are compound. This eyepiece may be recognized by the color of the fringe around the inside edge of the diaphragm when daylight is viewed through the eyepiece. Ordinary eyepieces show a blue fringe, while compensating eyepieces show a yellow, orange or red fringe.
The chromatic difference of magnification common to all high-power objectives can be corrected by using a compensating eyepiece. This eyepiece not only corrects for the chromatic difference of magnification introduced by the objectives but is also designed to correct image curvature to some extent. The compensating eyepiece should not be used with low-power achromatic objectives, because color aberration may be introduced. With the Nikon CF Optical System, it is no longer necessary to use compensating eyepieces.
FILAR MICROMETER EYEPIECES
The use of a Filar Micrometer eyepiece is one of the most precise and accurate means of measurement. For occasional measurements the eyepiece reticle serves a useful purpose; but for greater precision and accuracy, the filar eyepiece is absolutely essential.
The Filar Micrometer eyepiece is fitted with a drum which activates a vertical cross hair which travels across a fixed vernier scale. Some filar eyepieces have a graduated scale incorporated with the cross hair and both move simultaneously.
Since this filar micrometer eyepiece is used with objectives with a stated nominal value, it is important to always calibrate this filar micrometer with a stage micrometer.
The Nikon Metric Filar Micrometer eyepiece has a unique feature which makes calibration with different objectives relatively easy. The lower portion of the eyepiece has a graduated ring which when rotated optically changes the effective tube length so that the stage micrometer can coincide exactly with a portion of the internal scale of the filar. With most other filar micrometers, the chances of exactly super imposing the filar scale to the stage micrometer would depend on the accuracy of the objective lens.
MOUNTING A RETICLE IN THE EYEPIECE
The correct position of the reticle in the eyepiece is easily found, since in most oculars, a diaphragm or field stop is located in the ocular focal plane. To insert a reticle, unscrew the bottom retainer of the CFW 10x or CFW 15x eyepiece. Insert a 21mm reticle. The reticle will seat in its proper position. Care should be given to orientation of the reticle so that the numbers of the reticle can be seen in their correct position when viewed through the eyepieces. Reinsert the retaining ring and tighten. The reticle can be focused in the eyepiece by rotating the top of the eyepiece in the normal manner.
A simple reticle holder is available for mounting reticles in the stereo microscope eyepieces. A 24mm reticle is cemented into the reticle holder. The entire assembly is then inserted from the bottom of the eyepiece barrel and moved towards the eye lens until proper focus is obtained. The reticle holder will maintain its position due to spring tension of the holder on the sides of the eyepiece barrel. The focus point of the reticle can easily be changed to accommodate the user’s eye by moving the entire assembly in or out.
Before using this reticle holder, the diaphragm of the stereo eyepiece must be removed to allow the reticle holder to slip into the eyepiece tube.
Reticles used with measuring microscopes are supplied in a small reticle holder which easily drops into the optical head just below the eyepiece. These eyepieces have a diopter adjustment so that a clear image of the reticle can always be obtained. Measuring microscopes require a 24mm reticle.
Nikon Research Microscopes require a 21mm diameter reticle.
Nikon Stereo Microscopes require a 24mm diameter reticle.
Nikon Measurescopes require a 24mm diameter reticle. |